Medtronic Pacemaker Lawsuit Defective Medtronic Pacemaker
 
Ennis & Ennis, P.A. is a national law firm that specializes in mass torts, defective medical devices and product liablity cases. If you have been injured you need to know your legal rights. Contact the lawyers of Ennis & Ennis today for a free case evaluation.
Ennis & Ennis, P.A. News
Free Case Evaluation

Medtronic Warns Doctors About Defective Pacemakers

Medtronic Kappa and Sigma Series Pacemakers may have a defect which causes wire seperation and may lead to serious side effects for patients and may even result in death.

 


05/21/2009 - 
Medtronic recently sent out a "Dear Doctor" letter which warned healthcare providers of defective Medtronic pacemakers. Specifically the "Dear Doctor" letter was warning of defective Medtronic Kappa Series and defective Medtronic Sigma Series pacemakers. According to the information that was sent out to healthcare providers the Kappa and Sigma pacemakers may fail due to seperation of wires that connect the electronic circuit to other components inside the pacemakers. These defective pacemakers may pose a serious health risk for patients and may even lead to death. This new information about defective Kappa pacemakers and defective Sigma pacemakers is in addition to information that Medtronic sent out in 2005 regarding defective pacemakers with a battery depletion problem.

Since 1997, there have been over 1.7 million Kappa and Sigma devices implanted worldwide. Combining the new subsets of defective Kappa and Sigma pacemakers subject to the advisory with the Sigma pacemakers from the 2005 advisory brings the total number of active defective Medtronic devices up to 36,900 (~2% of all Kappa/Sigma devices implanted). Below is the actual "Dear Doctor" letter that Medtronic sent out in regards to the defective Kappa and Sigma series pacemakers.

May 2009

Dear Doctor,

We are writing to advise you about an issue with specific subsets of Kappa(r) and Sigma(r) series pacemakers that may fail at a higher than expected rate due to separation of wires that connect the electronic circuit to other pacemaker components (e.g., battery, connector). This may present clinically as loss of rate response, premature battery depletion, loss of telemetry, or no output. We are also updating the performance and patient management recommendations of a different subset of Sigma devices with the same possible clinical presentation, previously reported in a November 2005 advisory. Medtronic is communicating this information to the appropriate regulatory agencies.

Since 1997, there have been over 1.7 million Kappa and Sigma devices implanted worldwide. Combining the new subsets of Kappa and Sigma pacemakers subject to this advisory with the Sigma pacemakers from the 2005 advisory brings the total number of active devices now affected to 36,900 (~2% of all Kappa/Sigma devices implanted). 

Some patients, whose devices experience a wire separation resulting in a loss of pacing output, will experience a return of bradycardia symptoms (e.g. fainting or lightheadedness). In rare cases involving pacemaker dependent patients, loss of pacing output may result in death or serious injury. Medtronic has received two reports of patient death where it is possible but unclear whether this issue may have been a factor.

New Kappa and Sigma Issue

Worldwide, an estimated 15,200 active Kappa devices and 6,100 active Sigma devices, manufactured primarily between November 2000 and November 2002, are affected by this issue. Most of these devices have been implanted in patients for five years or longer and may be nearing normal elective replacement time.

Medtronic has observed 285 Kappa devices and 131 Sigma devices with this failure mechanism from these new Kappa and new Sigma device subsets. This represents 0.49% (Kappa) and 0.88% (Sigma) of the original affected implant population. Our modeling predicts failure rates of 1.1% (Kappa) and 4.8% (Sigma) over the remaining lifetime of these pacemakers due to this issue. There is no provocative testing that can predict which specific devices may fail, and no device programming can mitigate this issue if it occurs.

Performance Update to 2005 Sigma Advisory In November 2005, Medtronic

issued an advisory regarding a different subset of Sigma pacemakers. This advisory was related to wire separations caused by a particular cleaning solvent used in manufacturing and is not related to the current Kappa/Sigma wire separation issue. There are currently an estimated 15,600 active implants from this 2005 Sigma device subset.

Our original modeling predicted a failure rate from 0.17% to 0.30% over the remaining lifetime of these pacemakers and our advisory recommended that physicians determine whether device replacement was warranted. However, updated modeling predicts a failure rate of 3.9% over the remaining device life, and revised patient management recommendations are provided below.

Summary of Affected Devices

Product

Estimated # active devices in subset

Observed failure rate

Predicted lifetime failure rate

Estimated Average Remaining Life

New Kappa Subset

15,200

0.49%

1.1%

1.2 years

New Sigma Subset

6,100

0.88%

4.8%

3.8 years

2005 Sigma Subset

15,600

0.55%

3.9%

3.3 years

Total

36,900

Patient Management Recommendations

We realize that each patient requires unique clinical consideration and we support your judgment in caring for your patients. After consultation with Medtronic's Independent Physician Quality Panel, Medtronic offers the following recommendations for patients in the new Kappa, new Sigma, and 2005 Sigma subsets referenced in the table above:

* Physicians should advise their patients to seek medical attention immediately if they experience symptoms (e.g., fainting or lightheadedness).

* Physicians should consider device replacement for patients who are both pacemaker dependent and who have been implanted with a device in the affected subsets. Medtronic will offer a supplemental device warranty if the device is not already at elective replacement time.

* Physicians should continue routine follow up in accordance with standard practice for those patients who are not pacemaker dependent.

Consistent with the HRS recommendations on device advisory communications, we will be informing patients with registered devices in the subsets noted in the above table via letter dated May 27, 2009, advising them to contact you for more information. Medtronic's Patient Services group at 800-551-5544 is available to assist patients.

Physician and Patient Support

Attached are the specific model and serial numbers referenced in the table above of affected devices you are following according to our device registration records. You may also look up specific serial numbers online to determine if they are affected at KappaSigmaSNList.medtronic.com.

We will continue to monitor and analyze failure rates and to provide regular updates on the ongoing actual performance of these subsets and all Kappa and Sigma devices in our Product Performance Report, available at www.medtronic.com/crm/performance.

Continued Vigilance

There is an additional subset of Kappa devices where we have observed a much lower rate of occurrence of this issue. Approximately 96,000 devices of this subset remain active. We have observed a failure rate of approximately 0.04% in this subset and our modeling predicts a failure rate of 0.12% over the remaining device life. After review with our Independent Physician Quality Panel, we do not recommend any specific actions for this group of devices. We will continue to monitor performance and inform you if any specific patient management recommendations are warranted.

We regret any difficulties this may cause you and your patients. If you have any questions, or if we can be of assistance, please contact your local Medtronic Representative or Medtronic Technical Services at 800-505-4636.

Sincerely,

Tim Samsel

Vice President, Quality and Regulatory

Medtronic Cardiac Rhythm Disease Management


If you or a loved one have been injured as a result of defective Medtronic pacemaker with wire seperation, Medtronic defective battery, or had the device replaced or plan to have it replaced you may be entitled to compensation regardless if you suffered as a result of complications from surgery or the device itself. Contact the Medtronic attorneys of Ennis & Ennis today for a free confidential case evaluation. Our on staff nurse and Medtronic lawyers are standing by to answer any questions you may have regarding the Medtronic pacemaker defects, a possible Medtronic class action lawsuit, or any other type of Medtronic Pacemaker litigation.

Required fields in yellow
 
Name:
Email:
 
Telephone: (123-123-1234)
 
Mailing Address:
 
City:
 
State:
 
Zip Code:
On behalf of whom are you inquiring?
Is the person deceased?
Yes No
If the person is deceased please list the cause of death as listed on the death certificate:
 
Please list the date that the pacemaker was inserted (MM/DD/YYYY):
 
If applicable, Please list the date replaced (MM/DD/YYYY):
 
Please list the model originally inserted, and the new one inserted if applicable:
 
Were you originally contacted by your doctor regarding a possible defect in your pacemaker?
 
Please list any additional information you feel may be relevant to your potential Medtronic Pacemaker case:
 
How did you hear about us?
 
 
 

 

 

Nationwide Free Case Evaluations: 1-800-856-6405